MODEL FCL-1056

SECTION 5.0

 

PROGRAMMING THE ANALYZER

TABLE 5-1. DEFAULT SETTINGS (continued)

ITEM

 

CHOICES

DEFAULT

 

 

 

Measurement

 

 

 

 

 

 

1.

Chlorine (sensor 1)

 

 

 

 

 

 

 

 

a.

units

ppm or mg/L

ppm

 

 

 

 

 

 

b.

resolution

0.01 or 0.001

0.001

 

 

 

 

 

 

c.

input filter

0 to 999 sec

5 sec

 

 

 

 

2.

pH (sensor 2)

 

 

 

 

 

 

 

 

a.

pre-amplifier location

analyzer or sensor/junction box

analyzer

 

 

 

 

 

 

b.

solution temperature correction

on or off

off

 

 

 

 

 

 

c.

resolution

0.01 or 0.1

0.01

 

 

 

 

 

 

d.

input filter

0 to 999 sec

2 sec

 

 

 

 

 

 

e.

reference impedance

low or high

low

 

 

 

Temperature related settings

 

 

 

 

 

 

1.

Units

ºC or ºF

ºC

 

 

 

 

2.

Temperature compensation

automatic or manual

automatic

 

 

 

Security Code

 

 

 

 

 

 

1.

Calibrate/Hold

000 to 999

000

 

 

 

 

2.

Program/Display

000 to 999

000

 

 

 

pH Sensor Diagnostic Limits

 

 

 

 

 

 

1.

Reference offset

0 to 9999 mV

60 mV

 

 

 

 

2.

Diagnostics

on or off

on

 

 

 

 

3.

Glass impedance temperature

on or off

on

 

correction

 

 

 

 

 

 

 

4.

Glass fault (low impedance)

0 to 9999 MΩ

10 MΩ

 

 

 

 

5.

Glass fault (high impedance)

0 to 9999 MΩ

1500 MΩ

 

 

 

 

6.

Reference fault (high impedance)

0 to 9999 kΩ

40 kΩ

 

 

 

Calibration–pH

 

 

 

 

 

 

1.

Stabilization criteria

 

 

 

 

 

 

 

 

a.

time interval

0 to 99 sec

10 sec

 

 

 

 

 

 

b.

pH change

0.01 to 1.00 pH

0.02 pH

 

 

 

 

2.

User-entered slope

0.00 to 99.99 mV/pH

59.16 mV/pH

 

 

 

 

3.

User-entered offset

-999 to +999 mV

0 mV

 

 

 

Calibration–Analog Outputs

 

 

 

 

 

 

1.

4 mA

0.000 to 22.000 mA

4.000 mA

 

 

 

 

2.

20 mA

0.000 to 22.000 mA

20.000 mA

 

 

 

 

 

21

Page 31
Image 31
Emerson PN 51-FCL-1056 instruction manual Choices Default