MODEL FCL-1056

SECTION 5.0

 

PROGRAMMING THE ANALYZER

5.4.3Procedure – Configuring Alarms and Assigning Setpoints

1.Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER.

Program

Outputs

Alarms

Measurement

Temperature

2.Choose Alarms.

Alarms

Configure/Setpoint

Simulate

Synch Timers: Yes

3.Choose Configure/Setpoint.

Configure/Setpoint

Alarm 1

Alarm 2

Alarm 3

Alarm 4

4.Choose Alarm 1, Alarm 2, Alarm 3, or Alarm 4.

Alarm 1 Settings Setpoint: 0.000 ppm Assign: S1 Measure

Logic: Low

Deadband: 0.000 ppm

5.The screen summarizes the present configuration and setpoints. There are nine items: Setpoint, Assign (S1 is sensor 1 and S2 is sensor 2), Logic, Deadband, Interval time, On time, Recover time, and Hold while active. The last four items describe the operation of the timer. Only four items are shown at a time. To view the remaining items, scroll to the bottom of the screen and continue scrolling.

To make a change, move the cursor to the desired line and press ENTER. A screen will appear in which the present setting can be edited. Press ENTER to store the setting.

For an explanation of terms, see sections 5.4.1 and 5.4.2.

6. To return to the main display, press MENU then EXIT.

27

Page 37
Image 37