MODEL FCL-1056

SECTION 8.0

 

MAINTENANCE

SECTION 8.0

MAINTENANCE

8.1 ANALYZER

The analyzer used with the FCL needs little routine maintenance.

Clean the analyzer case and front panel by wiping with a clean soft cloth dampened with water ONLY. Do not use solvents, like alcohol, that might cause a buildup of static charge.

Sensor circuit boards are replaceable.

PN

24207-00 pH/ORP/ISE sensor board

24203-01 chlorine sensor board

To replace a board

WARNING

RISK OF ELECTRICAL SHOCK

Disconnect main power and relay contacts wired to separate power source before servicing

1.Turn off power to the analyzer.

2.Loosen the four screws holding the front panel in place and let the front panel drop down.

3.Loosen the gland fitting and carefully push the sensor cable up through the fitting as you pull out the circuit board.

4.Once you have access to the terminal strip, disconnect the sensor.

5.Unplug the sensor board from the main board. See Figure 3-2.

6.Slide the replacement board partially into the board slot. Plug the sensor board into the main board and reattach the sensor wires.

7.Carefully pull the sensor cable through the gland fitting as you push the sensor board back into the enclosure. Tighten the table glands.

8.Close the front panel.

9.Turn on power.

53

Page 63
Image 63
Emerson PN 51-FCL-1056 instruction manual Section Maintenance, Analyzer