MODEL FCL-1056

SECTION 5.0

 

PROGRAMMING THE ANALYZER

5.3.3. Procedure – Configure Outputs.

Program

Outputs

Alarms

Measurement

Temperature

1.Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER.

2.The cursor will be on Outputs. Press ENTER.

Outputs

Range

Configure

Simulate

Output 1 Configure

ProgramOutput 1

Output 2

Output 1 Configure

Assign:

 

S1 Meas

Range:

4-20 mA

Scale:

Linear

Dampening

0 sec

3.Choose Configure.

4.Choose Output 1 or Output 2.

5.The screen shows the present configuration. There are six items: Assign (S1 is sensor 1, S2 is sensor 2), Range, Scale, Dampening, Fault Mode, and Fault Value To display the fifth and sixth items, scroll to the bottom of the screen and continue scrolling.

To make a change, move the cursor to the desired line and press ENTER. A screen will appear in which the present setting can be edited. Press ENTER to store the setting.

For an explanation of terms, see sections 5.3.1 and 5.3.2.

6. To return to the main display, press MENU then EXIT.