SECTION 5.0

 

PROGRAMMING THE ANALYZER

5.6.4Procedure – Configuring Temperature Related Settings

1.Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER.

Program

Outputs

Alarms

Measurement

Temperature

2.Choose Temperature.

Temperature

Units:Units:

 

C

S1 Temp Comp:

Auto

S2 Temp Comp:

Auto

3.The screen summarizes the present sensor configuration. There will be between three and five items. Units, S1 Temp Comp, and S2 Temp Comp always appear. If manual temperature compensation was selected, the manual temperature values entered for each sensor (S1 and S2 Manual) will also appear.

To make a change, move the cursor to the desired line and press ENTER. A screen will appear in which the present setting can be edited. To store a setting, press ENTER.

4.For an explanation of terms, see sections 5.6.2 and 5.6.3.

5.To return to the main display, press MENU then EXIT.

5.7CONFIGURING SECURITY SETTINGS

5.7.1 Purpose

This section describes how to set security codes. There are three levels of security.

a.A user can view the default display and diagnostic screens only.

b.A user has access to the calibration and hold menus only.

c.A user has access to all menus.

The security code is a three digit number. The table shows what happens when different security codes (XXX and YYY) are assigned to Calibration/Hold and All. 000 means no security.

Calibration/Hold

All

What happens…

 

 

 

000

XXX

User enters XXX and has access to all menus.

 

 

 

XXX

YYY

User enters XXX and has access to Calibration and Hold

menus only. User enters YYY and has access to all menus.

 

 

 

 

 

XXX

000

User needs no security code to have access to all menus.

 

 

 

000

000

User needs no security code to have access to all menus.

 

 

 

33

Page 43
Image 43
Emerson PN 51-FCL-1056 Program Outputs Alarms Measurement, Temperature, S1 Temp Comp Auto S2 Temp Comp