MODEL FCL-1056

SECTION 6.0

 

CALIBRATION

6.4.4 Procedure-Manual Calibration

1.Obtain two buffer solutions. Ideally, the buffer pH values should bracket the range of pH values to be measured.

2.Remove the sensor from the flow cell. If the process and buffer temperatures are appreciably different, place the sensor in a container of tap water at the buffer temperature. Do not start the calibration until the sensor has reached the buffer temperature.

Calibrate?

Sensor 1

Sensor 2

Output 1

Output 2

3.Press MENU. The main menu screen appears. The cursor will be on Calibrate. Press ENTER.

4.Choose the sensor you wish to calibrate. Sensor 1 is the chlorine sensor. Sensor 2 is the pH sensor.

S2 Calibration

pH

ORP

Redox

Temperature

S2 pH Cal

ZeroBufferCalCal

Standardize

Slope:

56.19 mV/pH

Offset:

2 mV

5. Choose pH.

6.Choose Buffer Cal

S2 pH Buffer Cal Auto

Manual

S2 pH Manual Cal

Buffer 1

Buffer 2

7. Choose Manual.

8.Choose Buffer 1.

9.Rinse the sensor with water and place it in the first buffer. Be sure the glass bulb and reference junction are completely submerged. Swirl the sensor.

45

Page 55
Image 55
Emerson PN 51-FCL-1056 instruction manual Manual, Buffer