MODEL FCL-1056

3.1.3

Alarm wiring.

WARNING

Exposure to some chemicals may degrade the sealing properties used in the following devices: Zettler Relays (K1-K4) PN AZ8-1CH12DSEA

The alarm relay terminal strip is located just below the power connector on the power supply board. See Figure 3-2.

Keep alarm relay wiring separate from signal wiring. Do not run signal and power or relay wiring in the same conduit or close together in a cable tray.

SECTION 3.0

WIRING

FIGURE 3-2. Alarm relay connections.

3.2 SENSOR WIRING

The Model FCL is provided with sensor cables pre-wired to the analyzer. If it is necessary to replace the sensor cable, refer to the instructions below.

1.

Shut off power to the analyzer.

 

 

 

2.

Loosen the four screws holding the front panel in place and let it drop down.

3.

Locate the appropriate signal board.

 

 

 

 

 

 

 

 

 

 

 

 

Slot 1 (left)

 

Slot 2 (center)

Slot 3 (right)

 

 

 

 

 

 

 

 

 

 

communication

 

input 1 (chlorine)

input 2 (pH)

 

 

 

 

 

 

 

4.

Loosen the gland fitting and carefully push the sensor cable up through the fitting as you pull the board

 

forward to gain access to the wires and terminal screws. Disconnect the wires and remove the cable.

5.

Insert the new cable through the gland and pull the cable through the cable slot.

6.

Wire the sensor to the signal board. Refer to the wiring diagrams in Figures 3-3 and 3-4.

7.

Once the cable has been connected to the board, slide the board fully into the enclosure while taking up the

 

excess cable through the cable gland. Tighten the gland nut to secure the cable and ensure a sealed enclosure.

10

Page 20
Image 20
Emerson PN 51-FCL-1056 instruction manual Sensor Wiring, Alarm relay connections