MODEL FCL-1056

SECTION 5.0

 

PROGRAMMING THE ANALYZER

5.3.5 Procedure – Simulating Outputs

Program

Outputs

Alarms

Measurement

Temperature

1.Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER.

2.The cursor will be on Outputs. Press ENTER.

Outputs

RangeSimulate

Configure

Simulate

Simulate

ProgramOutput 1

Output 2

3.Choose Simulate.

4.Choose Output 1 or Output 2.

Output 1 Hold at

1 2.00 mA

5.Enter the desired simulated output current. To end the simulated current, press MENU or EXIT.

5.4 CONFIGURING ALARMS AND ASSIGNING SETPOINTS.

5.4.1 Purpose

The Model FCL analyzer has an optional alarm relay board. This section describes how to configure and assign setpoints to the alarm relays, simulate alarm action, and synchronize interval timers. CONFIGURE THE ALARMS FIRST.

1.Configuring an alarm means…

a. Assigning a sensor and measurement (chlorine, pH, or temperature) to an alarm. An alarm relay can also be used as a timer.

b.Selecting high or low logic. c. Choosing a deadband.

d.Setting the interval timer parameters.

2.Simulating an alarm means making the analyzer energize or de-energize an alarm relay.