MODEL FCL-1056

SECTION 6.0

 

CALIBRATION

6.4.6 Procedure-Entering a Known Slope and Offset

1.Calibration data, i.e., slope and offset at pH 7, can be entered directly into the analyzer if the data for the sensor are already known.

Calibrate?

Sensor 1

Sensor 2

Output 1

Output 2

2.Press MENU. The main menu screen appears. The cursor will be on Calibrate. Press ENTER.

3.Choose the sensor you wish to calibrate. Sensor 1 is the chlorine sensor. Sensor 2 is the pH sensor.

S2 Calibration

pHpH

ORP

Redox Temperature

S2 pH Cal

ZeroBufferCalCal

Standardize

Slope:

56.19 mV/pH

Offset:

2 mV

4. Choose pH.

5.Choose Slope or Offset. Enter the known slope and offset values. Always enter the slope as a positive number. Press ENTER to store the number. See Section 6.4.2 for more information.

6.To return to the main display, press MENU then EXIT.

48

Page 58
Image 58