SECTION 5.0

 

PROGRAMMING THE ANALYZER

5.9RESETTING THE ANALYZER

5.9.1 Purpose

This section describes how to clear user-entered values and restore default settings. There are three resets:

1.Resetting to factory default values clears ALL user entered settings, including sensor and analog output calibration, and returns ALL settings and calibration values to the factory defaults.

2.Resetting a sensor calibration to the default values clears user-entered calibration data for the selected sensor but leaves all other user-entered data unaffected.

3.Resetting the analog output calibration clears only the user-entered analog output calibration. It leaves all other user-entered settings unchanged.

5.9.2Procedure – Resetting the Analyzer

1.Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER.

Program

Temperature

Security

Diagnostics

Reset Analyzer

2.Scroll to the bottom of the screen and continue scrolling until Reset Analyzer is highlighted. Press ENTER.

Reset Analyzer

FactoryFactoryDefaultsDefaults

Sensor Cal Only

Output Cal Only

3.Choose whether to reset all user-entered values (Factory Defaults), sensor calibration (Sensor Cal Only), or output calibration (Output Cal Only). If you choose Sensor Cal Only or Output Cal Only a second screen appears in which you can select which sensor or output calibration to reset.

4. To return to the main display, press MENU then EXIT.