SECTION 5.0

 

PROGRAMMING THE ANALYZER

5.3.4. Procedure – Ranging Outputs.

Program

Outputs

Alarms

Measurement

Temperature

1.Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER.

2.The cursor will be on Outputs. Press ENTER.

Outputs

Range

Configure

Simulate

Output Configure

ProgramOutput 1

Output 2

Output Range

O1 S1 4mA 0.000 ppm

O1 S1 20mA: 10.00 ppm

O2 S1 4mA: 0.0C

O2 S1 20mA: 100.0C

3. Choose Range.

4.Choose Output 1 or Output 2.

5.The screen shows the present settings for the outputs. O1 is output 1, O2 is output 2, S1 is sensor 1, and S2 is sensor 2.

To make a change, move the cursor to the desired line and press ENTER. A screen will appear in which the present setting can be edited. Press ENTER to store the setting.

For an explanation of terms, see sections 5.3.1 and 5.3.2.

6. To return to the main display, press MENU then EXIT.