MODEL FCL-1056

S2 pH Auto Cal StartStartAutoAutoCalCal

Setup

S2 pH Auto Cal

Place Sensor in

Buffer 1

Press Enter

S2 pH Auto Cal

07.01 pH

S2 pH Auto Cal

Place Sensor in

Buffer 2

Press Enter

S2 pH Auto Cal

10.01 pH

S2 pH Auto Cal

Slope:

59.16 mV/pH

Offset:

10 mV

S2 Setup

Stable Time:

10 sec

Stable Delta:

0.02 pH

Buffer:

Standard

SECTION 6.0

CALIBRATION

8.Choose Start Auto Cal.

If you wish to change the stability criteria or the pH buffer list from the default values, choose Setup instead and go to step 14. The default stability is defined as a less than 0.02 pH change in 10 seconds. The default buffer list is Standard. See the table in section 6.4.2.

9.Rinse the sensor with water and place it in the first buffer. Be sure the glass bulb and reference junction are completely submerged. Swirl the sensor. Press ENTER.

10.Once the pH reading meets the stability requirements, the screen changes to show the nominal pH of the buffer. The nominal pH is the pH value at 25ºC. The displayed value is not correct, press the up or down arrow key until the correct value is showing. Press ENTER.

11.Remove the sensor from the first buffer. Rinse with water and place it in the second buffer. Be sure the glass bulb and reference junction are completely submerged. Swirl the sensor. Press ENTER

12.Once the pH reading meets the stability requirements, the screen changes to show the nominal pH of the buffer. If the displayed value is not correct, press the up or down arrow keys until the correct value is showing. Press ENTER.

13.If the calibration is successful, the screen at left will be displayed for five seconds. The display will then return to the screen in step 6.

If the calibration is not successful, the existing calibration data will not be changed. A screen will appear identifying the error (high slope, low slope, or offset error). For troubleshooting see section 9.6.

14.If you chose Setup in step 8, the screen at left appears. To make a change, move the cursor to the desired line and press ENTER. A screen will appear in which the present setting can be edited. Press ENTER to store the change.

15.To return to the main display, press MENU then EXIT.