n. b.:

Leave the protective cover on until the lamp is inserted. Avoid making fingermarks, or wipe off immediately.

Remove the defect lamp.

Put a new 12 V 100 W halogen lamp into the lamp holder without tilting (86.2 or 87.2).

Move the collector back.

Put on the lid and secure with screw (85.1 or 87.1).

Align the lamphousing against the microscope and secure with the clamp screw.

Connect the lamphousing to the power unit.

Switch off the microscope and the power unit.

Disconnect the appliance cable of the micro- scope and the power unit.

Slacken the clamp screw on the microscope and remove the lamphousing.

Slacken the screws (88.4 and 88.9) on the lid with a cross-tip screwdriver.

Pull the cut-out plug slightly out of the socket (88.11) and flip up lid.

n. b.:

Leave the protective cover on until the lamp is inserted. Avoid making fingermarks, or wipe off immediately.

Replacing the 12 V 100 W halogen lamp in lamphousing 106 z*

n. b.:

Always disconnect external transformers and the microscope from the mains when carrying out assembly work!

Slacken the fixing screws (88.10) on the lamp holder and pull out the lamp holder (Fig. 89).

Remove the defect lamp.

Put a new 12 V/100 W lamp into the lamp holder.

Push in the lamp holder and secure it with the screws (88.10).

Push the cut-out plug into the socket (88.11).

Close the lid and tighten the screws (88.4 and 88.9) on the lid.

Align the lamphousing against the microscope and secure with the clamp screw.

Connect the lamphousing to the power unit.