To correct defective eyesight:

Look through the right-hand eyepiece tube with your right eye and sharply focus the specimen with the fine drive.

Then look at the same area of the specimen with your left eye and rotate the left eyepiece tube until you obtain a sharp image. Do not use the fine drive for this.

If using eyepieces with adjustable eyelenses, do not compensate for defective eyesight by adjusting the eyepiece tube, but by adjusting the eyelens of the eyepiece.

Fig. 58

HCI BV22, ergo binocular tube with 15° – 50° viewing angle, field of view no. up to 22, eyepiece diameter 30 mm for HC PLAN 10x/20 or 22 eyepieces, interpupillary distance set- ting: 55 – 75 mm

1 Clamp screw, 2 Tube port, 3 Siedentopf binocular part

Trinocular tube HCI 3T22

Set the beamsplitter at visual observation by pushing in the switch rod. The switching positions are indicated by symbols on the

side of the tube.

100Ê % vis

– switch rod

50Ê % – 50Ê %

– switch rod

100Ê %Ê –Ê photo – switch rod

The eyepieces are set in exactly the same way as on the binocular tube.

Compensate defective eyesight by adjusting the eyelens of the eyepiece.

Fig. 59 HCI 3T22, trinocular tube with 45° viewing angle

Light path: 100 % vis

– switch rod

50 % – 50 %

– switch rod

100 % – photo – switch rod

Field of view no. up to 22, eyepiece diameter 30 mm for HC PLAN 10x/20 or 22 eyepieces, interpupillary distance set- ting: 55 – 75 mm

1 Clamp screw, 2 Tube port, 3 Siedentopf binocular part, 4 Photo/TV port, 5 Switch rod

2

 

 

 

2

 

 

 

3

 

 

 

4

 

 

 

 

 

 

 

3

 

 

 

 

 

1

 

5

1

 

68

Page 68
Image 68
Leica DM IRB manual Trinocular tube HCI 3T22, Side of the tube