Front controls

There is an LC display and five control keys on the front of the microscope.

lower Z position set

Z position

stepwidth

focus position set

focus keys on the right side of the microscope

-1 . 86mm S 1

10 0 xPH3 HH

D

 

 

 

 

 

 

 

 

 

Objective data

current

fluo cube

 

 

 

 

IC prism or

operating mode dry/imm

 

 

brightfield

 

 

 

 

LC display

The display gives information on the following functions:

Z position in μm or mm.

Set stepwidth for the fine focusing (S0, S1, S2, S3 and coarse focusing = SC, can be switched on and off by simultaneously pressing both focus keys).

Lower Z position set (symbol visible = threshold set).

Focus position set (symbol visible = threshold set).

Objective data (corresponding to the objec- tive engraving).

magnification

Phaco (PH1, PH2, . . . )

required objective IC prism (option).

Currently switched IC prism or brightfield position.

Fluo cube (option).

Operating mode (dry or immersion).

At the factory the microscope is set so that the fast focusing is controlled with the two keys on the right side of the microscope, indicated by the symbolon the right in the upper line of the display.

Similarly, if this symbol appears on the left side of the display, it means that the focusing is controlled with the two keys on the left of the microscope.

n. b.:

When the microscope is installed the influence of the rotation direction of the handwheel on the focusing direction of the objective nosepiece can be reversed (p. 48).