Assembling and exchanging incident light lamps

Assembling and exchanging Hg and Xe lamps

Power units:

Hg and Xe lamps are powered by separate power units.

Please make sure to read the special manuals for these power units.

Lamphousing 106 z L

Besides the halogen lamp, the following gas discharge lamps can be used, which each require different lamp holders (Fig. 35) and power units:

Type

 

Average life span

Hg ultra high pressure lamp

50

W (A.C.)

100 h

Xe high pressure lamp

75

W (D.C. stabilised)

400 h

Hg ultra high pressure lamp 100

W (D.C. stabilised/non-stabilised)

200 h

Hg ultra high pressure lamp 100

W (D.C. stabilised/non-stabilised,

 

 

 

type 103 W/2)

300 h

Fig. 35 Lamp holders for gas discharge lamps

1 Upper clamp, 2 Seal point of the burner, 3 Lower clamp, 4, 6 Lamp holder screws, 5 Socket for cut-out plug, 7 Protective cover

Hg 50

 

 

 

Xe 75

1

 

 

 

 

 

 

 

 

 

 

 

1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

7

 

 

 

 

 

 

 

 

 

 

2

 

 

 

 

 

 

 

 

 

 

 

3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4

 

 

 

 

 

 

 

3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Hg 100

1

3

Hg 100 Stab.

1

3

31