Technical data of the power unit

General technical data

For indoor use only

Mains voltage:

Frequency:

Power consumption:

Fuses:

Ambient temperature:

Relative humidity:

Overvoltage category:

Contamination class:

Technical specifications

Lamp voltage:

Voltage setting:

Maximum lamp voltage:

Soft start:

Mains voltage dependence UN = mains voltage

ULa = lamp voltage

U

N

: 90Ê –Ê 250

Vac, U= 12 V:

U

: 90Ê –Ê 250

La

N

Vac, U< = 11 V:

U

 

La

N

: 100Ê –Ê 130 Vac, U< = 11 V:

U

 

La

N

: 200Ê –Ê 250 Vac, U< = 11 V:

 

 

La

Lamp voltage drift 0 to 10 min.:

Efficiency:

Short-circuit and open-circuit proof

Life span:

90Ê –Ê 250 V~ 50Ê –Ê 60 Hz

160W

T 4 A

10Ê –Ê 36°C

–Ê 80 % up to 30°C II

Adjustable from

2.5 V ± 5 % to 12 VÊ –Ê 5 % /8.5 A

Potentiometer 5 KOhm

Rotated clockwise for maximum intensity

12.0 V in the range 90 V to 250 V~

Rise time up to

maximum output voltage 0.2 to 1 second

<– 5 %

<± 1 %

<± 0.5 %

<± 0.5 %

<2 %

approx. 75 %

> 50,000 hours

126

Page 126
Image 126
Leica DM IRB manual Technical data of the power unit General technical data, Technical specifications, ULa = lamp voltage