Chapter 8
8-28 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Operating Instructions
Patient Data Screen You can view the patient data screen (Figure 8-33) by pressing the PATIENT

DATA button in the top bar. The format of this scr een is the same in VCV, PCV,

and NPPV. Table 8-7: “Patient Data Definitions: Range, Units & Resolution”

summarizes the definitions, display ranges, and resolution for patient data

parameters.

Figure 8-33: Patient Data Screen

Patient Data Definitions: Range, Units & Resolution

Parameter Description Display Range Resolution
PIP Peak Inspiratory Pressure
The maximum airway pressure
during inspiration.
-20.0 to 130
cmH2O (hPa) 0.1 for -20.0 to
99.9
1 for 100 to 130
MAP Mean Airway Pressure
The average of airway pressure
during one full breath cycle.
-20.0 to 120
cmH2O (hPa) 0.1 for -20.0 to
99.9
1 for 100 to 130
Pe End Pressure at End Expiration
The pressure measured at the
end of expiration.
-20.0 to 99.9
cmH2O (hPa) 0.1
Pi End Pressure at End Inspiration
The pressure measured at the
end of inspiration (or at the end
of Insp Hold if it is >0).
-20.0 to 130
cmH2O (hPa) 0.1 for -20.0 to
99.9
1 for 100 to 130
Tidal Vol. Exhaled Tidal Volume
The volume exhaled at each
breath. Compliance
compensated (if enabled).
0 to 9999 ml
(BTPS) 1
Table 8-7: Patient Data Definitions: Range, Units & Resolution (Sheet 1 of 2)