Chapter 13
13-16 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Options and Accessories
3. Position the manifold support bracket so that the Wall O2 connection
of the manifold fits snugly into the manifold support bracket grommet.
Using the Allen wrench, attach the bracket to the cart using the two
M6 x 10mm screws.
Figure 13-11: Attaching Manifold Support Bracket
4. If you will be using O2 cylinders, connect either or both of the oxygen
hoses from the O2 cylinders to the left and right inlets of the oxygen
manifold assembly. These inlets are marked "O2 Cylinder".
Figure 13-12: Completed Manifold Assembly
Oxygen Manifold
Kit - P/N 1004415
For Use With Exprit
Ventilator Only
CAUTION
Use Oxygen Only
Operating Pressure Range:
50-90 PSIC at 200 SLPM
WALL CONNECTION
O
2
CYLINDER
O
2
CYLINDER
RESPIRONICS
®
Wall O2 connection