Chapter 9

Alarms

 

Alarm Alert Messages (Continued)

 

 

Alert Message

Description

 

 

Low Insp Pressure

If the peak airway pressure during any mandatory inhalation

 

is less than the Low Insp Pressure limit, the ventilator will

 

immediately signal a high priority alarm.

 

Note: If an inspiration is triggered immediately after the

 

mandatory minimum exhalation time of 200 msec, this

 

alarm will not be activated unless this condition occurs for

 

three consecutive breaths.

 

 

Low Minute Volume

Indicates the measured patient minute ventilation (VE) is

 

below the set limit. It is elevated to high urgency after one

 

minute.

 

Note: A Low Minute Volume setting of zero (0) will disable

 

this alarm.

 

 

Low O2

A high urgency alarm indicates that the monitored O2

 

concentration is 18% or at least 6% below set value for 30

 

seconds. Verify operation of the O2 sensor. Alarm cannot be

 

manually reset.

 

 

Low O2 Supply

A high urgency alarm indicates that O2 gas supply is below

 

acceptable levels and the %O2 setting is above 21%. Check

 

O2 gas connections and inlet filter. Alarm cannot be

 

manually reset.

 

 

Low PEEP

If PEEP is less than the operator set Low PEEP limit for one

 

second during exhalation, triggers a medium priority alarm.

 

If the Low PEEP alarm remains active for one minute, the

 

alarm moves to high priority.

 

Note: A Low PEEP setting of zero (0) will disable this alarm.

 

Note: If an inspiration is triggered immediately after the

 

mandatory minimum exhalation time of 200 msec, this

 

alarm will not be activated unless this condition occurs for

 

three consecutive breaths.

 

 

Low Tidal Volume

The ventilator triggers a medium urgency alarm to indicate

 

that the NPPV (mandatory or spontaneous breath) tidal

 

volume is less than the set limit. It is elevated to a high

 

urgency alarm after one minute.

 

Note: A Low Tidal Volume setting of zero (0) will disable this

 

alarm.

 

 

Table 9-1: Alarm Alert Messages (Sheet 3 of 4)

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

9-7