Chapter 13
13-66 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Options and Accessories
Dynamic Resistance
Dynamic resistance is calculated using this equation.
Where:
n = Number of samples taken during for the calculation
i = Sample number
Qi = Patient flow
Vi = Lung volume for the /th sample number
Pi = Pressure sensor measurement for the /th sample number
Static C and R
Static C and R are calculated using these equations:
Where:
RL
QiViPiViPiQiVi
i0=
n
i0=
n
i0=
n
i0=
n
QiVi
i0=
n
2
Qi
2
i0=
n
Vi
2
i0=
n
------------------------------------------------------------------------------------------------=
Qitd
=
CLV
Peplat P0
()
--------------------------------=
CRL
Pei Peplat PVplat CL
()()
QLend
------------------------------------------------------------------------=