Chapter 13

Options and Accessories

Information type

Description

Start MIP Press & Hold button

Starts the MIP maneuver. Changes to Continue when the maneuver is complete or when finger is lifted from the button. Changes to Start when Continue is touched.

Start P0.1 button

Starts the P0.1 maneuver. Changes to Stop when the maneuver is

 

started. Changes to Continue when the maneuver is complete or when

 

the Stop button is touched. Changes to Start when Continue is

 

touched.

 

 

MIP window

Displays the most recently accepted maneuver results. The date and

 

time of the maneuver appears below each window.

 

 

P0.1 window

Displays the most recently accepted maneuver results. The date and

 

time of the maneuver appears below each window.

Scale (button at left of graphic display)

Allows you to adjust graphic scales manually. The default scale for the MIP/P0.1 screen is –100 cmH2O to the High Pressure limit. Default scales are in effect every time you enter the MIP/P0.1 screen, and the graph reverts to default scales when the maneuver is complete.

Rescale button

Allows software to adjust the vertical scales for waveforms and vertical

 

and horizontal scales for loops. The graph cannot be rescaled during

 

the MIP maneuver.

 

 

 

Table 13-6: Summary of MIP/P0.1 Screen Information

Static C and R

Maneuver

The Static C and R screen (Figure 13-31) allows you to perform a static compliance and resistance maneuver and displays the maneuver results. The maneuvers are carried out on machine or operator initiated breaths. If the patient triggers a breath when a maneuver is scheduled, the ventilator does not perform the maneuver. Although the maneuver does not require patient effort, patient interference can affect the accuracy of maneuver results. The Static C and R maneuver is available in VCV only. It is not available in PCV and NPPV.

NOTE: Depending on the High Pressure limit setting, if the pre-maneuver flow pattern is ramp, delivering a square waveform during the Static C and R maneuver can trigger a High Inspiratory Pressure alarm and cancel the maneuver. Should this happen, adjust alarm settings as needed.

How to perform Static C and R maneuver

1.Enter the Static C and R screen: touch the Mechanics icon button, then Static C and R.

2.Touch the Start button. If you do nothing, the maneuver automatically ends at the next mandatory inspiration.

Once touched, the Start button changes to Stop. The Stop button allows you to cancel the maneuver at any time. A high-priority alarm automatically cancels the maneuver.

13-56

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A

Page 220
Image 220
Philips V200 manual How to perform Static C and R maneuver, 13-56, Information type

V200 specifications

The Philips V200 is a robust mobile device that combines functionality with sleek design, catering to users who prioritize efficiency and modern technology. This smartphone is engineered to provide a seamless user experience, integrating advanced features and specifications that enhance daily activities.

One of the standout features of the Philips V200 is its vibrant display. The device boasts a large, high-resolution screen that renders visuals with stunning clarity and color accuracy. This makes it an excellent choice for multimedia consumption, whether you're watching videos, playing games, or viewing photos. The touchscreen interface is smooth and responsive, allowing for effortless navigation across applications.

Under the hood, the Philips V200 is powered by a reliable processor, ensuring fast performance and multitasking capabilities. With ample RAM, users can run multiple applications simultaneously without experiencing lag, making it suitable for both work and leisure. This processing power is complemented by substantial internal storage, providing enough space for apps, photos, videos, and other essential files.

Connectivity is another highlight of the Philips V200. It supports various networks and is equipped with 4G LTE capabilities, ensuring users stay connected at high speeds wherever they go. Bluetooth and Wi-Fi functionalities also enhance connectivity options, allowing easy sharing of files and media between devices.

The camera system on the Philips V200 is designed to cater to photography enthusiasts. With a high-resolution rear camera and a secondary front camera, users can capture stunning images and selfies. The device also includes several camera features such as portrait mode and various filters, making photography more enjoyable and creative.

The Philips V200 places significant emphasis on battery life, with a long-lasting battery that ensures all-day usage. Users can rely on the device for extended periods without the need for frequent charging, which is particularly useful for those with busy lifestyles.

In terms of software, the Philips V200 runs on a user-friendly operating system that provides access to a plethora of applications and services. The user interface is intuitive, making it easy for users to customize their devices according to their preferences.

Overall, the Philips V200 stands out in the competitive smartphone market by offering a well-rounded package of features, performance, and design. It is a practical choice for anyone seeking a reliable and stylish mobile device that can handle both everyday tasks and more intensive applications with ease. Whether you are a professional, a student, or simply someone who enjoys technology, the Philips V200 is engineered to meet diverse needs.