Appendix A
Appendix A-8 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
RS-232 Communications Protocol
Vital Capacity Time of
Last Maneuver 11-14-
01◆◆13:11 1 minute 01-01-70
0:00 to
01-19-38
3:14
N/A Time of Last
Maneuver display
from the Vital
Capacity
Screen.15
character field, 24
hour clock: MM-
DD-YYuuHH:MM
MIP -54.10.1 for -99.9
to 99.9
1 for -100
and 100 to
200
-100 - 200 cmH20 MIP display from
MIP/P0.1 display
Screen
MIP Time of Last
Maneuver
11-14-01◆◆13:11 1 minute 01-01-7
0 0:00 to
01-19-38
3:14
N/A MIP Time of Last
Maneuver display
from MIP/P0.1
Screen15
character field, 24
hour clock: MM-
DD-YYuuHH:MM
P0.1 -2.3◆◆ 0.1for -99.9
to 99.9
1 for -100
and 100 to
200
-100 - 200 cmH20 P0.1 display from
MIP/P0.1 display
screen
P0.1 Time of Last
Maneuver
11-14-01◆◆13:11 1 minute 01-01-70
0:00 to
01-1-/38
3:14
N/A P0.1 Time of Last
Maneuver display
from MIP/P0.1
Screen.15
character field, 24
hour clock: MM-
DD-YYwwHH:MM
Static Resistance 5.43◆◆ 0,01 for
0.00 to 9.99
0.1 for 10.0
to 99.9
1 for 100 to
400
0.00 - 400 cmH20
/L/Sec Static Resistance
display from the
Static C & R screen
Static Compliance 19.2◆◆ 0.01 for
0.00 to 9.99
0.1 for 10.0
to 99.9
1 for 100 to
350
0.00 - 350 mL/
cmH2OStatic Compliance
display from the
Static C & R
Screen
Static C & R Time of
Last Maneuver
11-14-01◆◆13:11 1 minute 01-01-70
0:00 to
01-19-38
3:14
N/A Static C & R Time
of Last Maneuver
display from Static
C & R Screen.15
character field, 24
hour clock: MM-
DD-YYwwHH:MM
Ventilation Report (Continued)
Description Example Resolution Range Units Comments

Table A-1: Ventilation Report (Sheet 7 of 10)