Appendix A

RS-232 Communications Protocol

Non-Invasive Positive Pressure Ventilation Settings Report (Continued)

Description

Example

Resolution

Range

Units

Comments

NPPV High Leak

60◆◆◆◆

1

1 - 60

LPM

 

Alarm Limit

 

 

 

 

 

 

 

 

 

 

 

Stop Code

0x03

N/A

N/A

N/A

ASCII End

 

 

 

 

 

Transmission

 

 

 

 

 

Character (ETX)

 

 

 

 

 

 

Table A-4: Non-Invasive Positive Pressure Ventilation Settings Report (Sheet 3 of 3)

Patient Data Report (PTDT)

When the ventilator receives PTDT followed by a carriage return, it will respond by transmitting the information shown in Table A-5: “Patient Data Report”. The ventilator responds to the PTDT command by returning a string with a variable length. Fields 2 through 4 define the length of the message. The last character transmitted is a stop code indicating the end of the message. The second field indicates the number of characters between the start and stop codes. The third field indicates the number of fields between the start and stop codes. The fourth field is the start code, 0x02. The last field in the string is the stop code, 0x03.

Patient Data Report

Description

Example

Resolution

Range

Units

Comments

 

 

 

 

 

 

Command Name

PTDT

N/A

N/A

N/A

 

 

 

 

 

 

 

Number of

275

N/A

N/A

N/A

3 character field

characters between

 

 

 

 

 

the start and stop

 

 

 

 

 

codes

 

 

 

 

 

 

 

 

 

 

 

Number of fields

32

N/A

N/A

N/A

2 character field

between the start

 

 

 

 

 

and stop codes

 

 

 

 

 

 

 

 

 

 

 

Start Code

0x02

N/A

N/A

N/A

ASCII Start

 

 

 

 

 

Transmission

 

 

 

 

 

Character (STX)

 

 

 

 

 

 

Time of request

13:45

N/A

N/A

N/A

24 hour clock,

 

 

 

 

 

hh:mm

 

 

 

 

 

 

Date

FEB231997

N/A

N/A

N/A

12 character field,

 

 

 

 

 

MMMDDYYYY

 

 

 

 

 

 

Measured Peak

24.1◆◆

0.1 for -20.0 to 99.9

-20.0 - 130

cmH2O

The PIP value from

Inspiratory

 

1 for 100 to 130

 

(hPa)

the Patient Data

Pressure

 

 

 

 

screen.

 

 

 

 

 

 

Measured Mean

5.6◆◆◆

0.1 for -20.0 to 99.9

-20.0 -120

cmH2O

The MAP value

Airway Pressure

 

1 for 100 to 120

 

(hPa)

from the Patient

 

 

 

 

 

Data screen.

 

 

 

 

 

 

Table A-5: Patient Data Report (Sheet 1 of 4)

Appendix A-18

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A

Page 304
Image 304
Philips V200 manual Patient Data Report Ptdt, Appendix A-18

V200 specifications

The Philips V200 is a robust mobile device that combines functionality with sleek design, catering to users who prioritize efficiency and modern technology. This smartphone is engineered to provide a seamless user experience, integrating advanced features and specifications that enhance daily activities.

One of the standout features of the Philips V200 is its vibrant display. The device boasts a large, high-resolution screen that renders visuals with stunning clarity and color accuracy. This makes it an excellent choice for multimedia consumption, whether you're watching videos, playing games, or viewing photos. The touchscreen interface is smooth and responsive, allowing for effortless navigation across applications.

Under the hood, the Philips V200 is powered by a reliable processor, ensuring fast performance and multitasking capabilities. With ample RAM, users can run multiple applications simultaneously without experiencing lag, making it suitable for both work and leisure. This processing power is complemented by substantial internal storage, providing enough space for apps, photos, videos, and other essential files.

Connectivity is another highlight of the Philips V200. It supports various networks and is equipped with 4G LTE capabilities, ensuring users stay connected at high speeds wherever they go. Bluetooth and Wi-Fi functionalities also enhance connectivity options, allowing easy sharing of files and media between devices.

The camera system on the Philips V200 is designed to cater to photography enthusiasts. With a high-resolution rear camera and a secondary front camera, users can capture stunning images and selfies. The device also includes several camera features such as portrait mode and various filters, making photography more enjoyable and creative.

The Philips V200 places significant emphasis on battery life, with a long-lasting battery that ensures all-day usage. Users can rely on the device for extended periods without the need for frequent charging, which is particularly useful for those with busy lifestyles.

In terms of software, the Philips V200 runs on a user-friendly operating system that provides access to a plethora of applications and services. The user interface is intuitive, making it easy for users to customize their devices according to their preferences.

Overall, the Philips V200 stands out in the competitive smartphone market by offering a well-rounded package of features, performance, and design. It is a practical choice for anyone seeking a reliable and stylish mobile device that can handle both everyday tasks and more intensive applications with ease. Whether you are a professional, a student, or simply someone who enjoys technology, the Philips V200 is engineered to meet diverse needs.