Chapter 13

Options and Accessories

Parameter

Description

 

 

Static compliance

Procedure: Patient participation is not required, and patient interference

and resistance

can affect the accuracy of the maneuver. If the patient initiates a breath

(static C and R)

when a maneuver breath is scheduled, the ventilator does not perform the

maneuver

maneuver. Available in VCV only. Not available in PCV and NPPV.

 

During the maneuver: The ventilator delivers a square waveform, regardless

 

of the current setting. No other settings change during the static C and R

 

maneuver. The maneuver is not performed unless a machine or operator

 

triggered breath is delivered within 20 seconds. The inspiration time must

 

be between 200 ms and 5 seconds for the maneuver to occur, and the

 

maximum plateau duration is 2.5 seconds.

 

Results display: Static compliance is displayed in mL/cmH2O. Static

 

resistance is displayed in cmH2O/L/s. All compliance and resistance values

 

are measured on time or operator triggered mandatory breaths only. During

 

the last saved Static C and R maneuver, Plateau Pressure (Pplat) is

 

measured and displayed as a static value. Static C and R are time-stamped

 

based on ventilator time and date. Static compliance values from 0 to 350

 

mL/cmH2O and static resistance values from 0 to 400 cmH2O/L/s are

 

displayed: If either value exceeds its limits or a stable plateau pressure

 

cannot be achieved, only a message (no data) is displayed.

 

Range:

 

Static compliance: 0 to 350 mL/cmH2O

 

Static resistance: 0 to 400 cmH2O/L/s

 

Plateau Pressure: -20 to 130 cmH2O

 

Resolution:

 

Static compliance: 1 mL/cmH2O

 

Static resistance: 1 cmH2O/L/s

 

Plateau Pressure: 1 cmH2O

 

Accuracy:

 

Static compliance: ± 1 mL/cmH2O ± 20% of true value

 

Static resistance: ± 3 cmH2O/L/s ± 20% of true value

 

Plateau Pressure: ± (0.5 cmH2O) ± (10%) of true value

Inspiratory time

Calculated inspiratory time divided by total breath time, or percent

divided by total

inspiratory time. An eight-breath running average. Available in VCV, PCV,

breath time (TI/

and NPPV.

TTOT).

 

 

Applicable to spontaneous and pressure-supported breaths only. If there

 

were no spontaneous or pressure-supported breaths within the last one

 

minute, - - - - is displayed.

 

 

Table 13-9: Respiratory Mechanics Maneuver Summary and Data Specifications

13-64

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A