Chapter 13
13-42 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Options and Accessories
VCV Apnea
Respiratory Rate
setting
12.0 0.1 or 1 1.0 – 9.0
10 – 80
BPM
Apnea Peak Flow
setting (same as Peak
Flow setting)
5 1 3 – 140 LPM Outputs VCV Peak
Flow setting
VCV Apnea Oxygen
Concentration setting
(same as Oxygen
Concentration
setting)
21 1 21 – 100 % Outputs Oxygen
Concentration
setting
Pressure Support
setting 0 10 100cm
H2OOutputs value from
active ventilation
type (VCV or PCV).
When active type
is NPPV then IPAP
– EPAP is used.
VCV Flow Pattern
setting SQUARE N/A SQUARE,
RAMP
N/A
Not used  N/A N/A N/A Outputs example
value.
Not used  N/A N/A N/A Outputs example
value.
100% Oxygen setting ON N/A ON
OFF
N/A
Not used  N/A N/A N/A Outputs example
value.
Not used  N/A N/A N/A Outputs example
value.
Not used  N/A N/A N/A Outputs example
value.
Measured total
respiratory rate 0.0 0.1 or 1 0.0 – 9.9
10 - 150
BPM For values greater
than 150 output
set to 150.
Measured tidal
volume 1.15 0.01 0.00 – 9.99 L Outputs 
for values out of
range.
Measured total
minute volume 0.00 0.0 1 or 1 0.00 – 9.99
10.0 – 99.9
L For values out of
range output set to
99.9.
Measured
spontaneous minute
volume
0.00 0.01 or 0.1 0.00 - 9.99
10.0 – 99.9 L Outputs 
when ventilation
mode is A/C. For
values out of range
output set to 99.9.
Ventilation Report
Description Example Resolution Range Units Comments

Table 13-4: Ventilation Report (Sheet 3 of 10)