Options and Accessories
REF 1057983 A Respironics V200 Ventilator Operator’s Manual 13-57
Chapter 13
3. At the end of a successful maneuver, a pop-up screen displays the
result of the maneuver and asks you to accept or reject the results by
touching the Accept or Reject button.
4. At the end of the maneuver (whether successful or cancelled), the
Stop button changes to Continue. While the Continue button is
displayed, the graph is frozen and can be rescaled. Touch Continue or
exit the screen to unfreeze the display.
5. Touch Continue: the button changes to Start and you can repeat the
maneuver.
6. Return the active ventilation mode from VCV to the previous mode if
necessary.
Figure 13-31: Static C and R Screen
Information type Description
Pressure--Volume or
Pressure--Time button Pressure--Volume: Displays a real-time pressure-volume loop.
Pressure--Time: Displays a real-time pressure-time waveform.
Start button Starts the maneuver. Changes to Stop when the maneuver is started.
Changes to Continue when the maneuver is complete or when the Stop
button is touched. Changes to Start when Continue is touched.
Static C and Static R
windows Displays the most recently accepted maneuver results. The date and
time of the maneuver appear below each window.
Table 13-7: Summary of Static C and R Screen Information
Time
Stamp
Rescale
button
Start
button
Static C and R Windows
Pressure-Volume and Pressure-Time buttons
P Plat
window
Scale
button