Chapter 13
13-36 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Options and Accessories
Analog Output (Chart
Recorder) The Communications-Print option enables the ventilator to send ventilator data
to a chart recorder or similar device using the analog output port.
Data displayed via the chart recorder is for reference purposes only. Decisions
for patient care should not be based solely on the data displayed via the chart
recorder.
Table 13-3 summarizes the Analog Output pinout.
X HIGH INT
TEMP High
Temperature Temperature within ventilator enclosure exceeds
maximum limit.
X HIGH INT
O2% High Internal
O2 Oxygen concentration within ventilator enclosure
exceeds maximum limit.
NOTE: The following alarm messages are not displayed on the CMS:
Bad
ADCWrap
Sensor
ERROR
Bad Bat
Volt
Sensor
Service Due
Bad Int
O2 Sensor Using Default Altitude
Bad Int
Temp
Sensor
Using Default Compl
Battery
Charging Using Default Settings
Available
on
CareNet
CMS
message V200
equivalent Comments
Respironics V200 Ventilator Analog Output Port Pinout
Pin Number Signal
1 Flow signal return
2 Volume signal return
3 Pressure signal return
4 Alarm signal return
5 Nebulizer signal return
9 Flow (0 V = -200 LPM, 5 V = 200 LPM)
10 Volume (0 V = 0 ml, 5 V = 2500 ml)
11 Pressure (0 V = -20 cmH2O, 5 V = 120 cmH2O)
12 Alarm (0 V = active alarm, 1.5 V = no alarm or silenced alarm)