Chapter 12
12-6 Respironics V200 Ventilator Operator’s Manual REF 1057983 A
Technical Specifications
Front Panel Keys
I:E 9.9:1 to 1:99 none 0.1 for 9.9:1-1:9.9;
1 for 1:10-1:99 ±10%
Peak Inspiration
Pressure -20.0 to 130 cmH2O
(hPa) 1 for 100-130;
0.1 for -20.0 to 99.9 ±10%
End Inspiration
Pressure (Pi End) -20.0 to 130 cmH2O
(hPa) 1 for 100-130;
0.1 for -20.0 to 99.9 ±10%
Mean Airway
Pressure -20.0 to 120 cmH2O
(hPa) 1 for 100-120;
0.1 for -20.0 to 99.9 ±10%
Delivered O2% 0.0 to 110 -- 1 for 100-110;
0.1 for -20.0 to 99.9 ±3 vol%
Total Respiratory Rate 0.0 to 150 Bpm 1 for 10-150;
0.1 for 0.0 to 9.9 ±10%
Spontaneous
Respiratory Rate 0.0 to 150 Bpm 1 for 10-150;
0.1 for 0.0 to 9.9 ±10%
End Exhalation
Pressure (Pe End) -20.0 to 99.9 cmH2O
(hPa) 0.1 ±10%
Patient Leak (Pt Leak) 0.0 to 140 LPM 0.1 for 0.0-99.9
1 for 100-140 Not
specifieda
Percent Patient Trigger
(%Pt Trigger) 0.0 to 99.9 % 0.1 for 0.0-99.9
1 for 100 ±10%
a. Due to the variable nature of patient leaks, this parameter is an estimate only.

Patient Data Range, Resolution, Units & Accuracy (Continued)

Patient Data Display Range Units Resolution Accuracy

Table 12-10: Patient Data Range, Resolution, Units & Accuracy (Sheet 2 of 2)

Front Panel Keys

Alarm Silence Screen Lock

Alarm Reset Expiratory Hold

100% O2Accept

Manual Breath

Table 12-11: Front Panel Keys