Chapter 13

Options and Accessories

Available

CMS

V200

 

on

Comments

message

equivalent

CareNet

 

 

 

 

 

 

 

 

X

HIGH

High Leak

High leak alarm limit violation. Yellow Alarm.

 

LEAK

Rate

 

 

RATE

 

 

 

 

 

 

X

LOW EXH

Low Vt

Low mandatory tidal volume limit violation. Yellow

 

MAND TV

Mandatory

Alarm.

 

 

 

 

X

HIGH EXH

High Minute

High exhaled minute volume limit violation. Yellow

 

MV

Volume

Alarm.

 

 

 

 

X

LOW EXH

Low Tidal

Low exhaled tidal volume limit violation. Yellow Alarm.

 

TV

Volume

 

 

 

 

 

X

LOW EXH

Low Vt

Low spontaneous tidal volume limit violation. Yellow

 

TV-SPNT

Spontaneous

Alarm.

 

 

 

 

X

INS TI

I-Time Too

Maximum inspiratory time exceeded. Yellow Alarm.

 

TOO

Long

 

 

LONG

 

 

 

 

 

 

X

HIGH

High

High respiratory rate limit violation. Yellow Alarm.

 

RESP

Respiratory

 

 

RATE

Rate

 

 

 

 

 

X

HIGH

High O2

High measured oxygen concentration. Yellow Alarm.

 

O2%

 

 

 

 

 

 

X

LOW

Low PEEP

Low PEEP limit violation. Yellow Alarm.

 

PEEP/

 

 

 

EPAP

 

 

 

 

 

 

X

LOW

Low EPAP

Low EPAP limit violation. Yellow Alarm.

 

PEEP/

 

 

 

EPAP

 

 

 

 

 

 

X

BACKUP

Battery In Use

Ventilator has switched from AC power to Backup

 

BATT ON

indicator on

Battery operation. Yellow Alarm.

 

 

ventilator

 

 

 

front panel

 

 

 

 

 

Inops

 

 

 

 

 

 

 

X

LOSS OF

Gas Supplies

Oxygen and air sources lost, safety valve open.

 

GAS-SVO

Lost-SVO

 

 

 

 

 

X

LOW

Low Backup

Low Backup Battery voltage.

 

BACKUP

Battery

 

 

BATT

 

 

 

 

 

 

X

ALARM

Audible Alarm

Audible alarm failure.

 

FAILURE

Failed

 

 

 

 

 

X

AIR

Air Source

Internal air source failure.

 

FAULT

Fault

 

 

 

 

 

X

O2 VALVE

O2 Valve

Defective oxygen valve.

 

STUCK

Stuck Closed

 

 

 

 

 

X

EXH

Exp Valve

Defective exhalation valve.

 

VALVE

Stuck Open

 

 

STUCK

 

 

 

 

 

 

REF 1057983 A