Chapter 3: Operation

Signal parameters on ILS display

Following signal parameters are indicated on the display:

 

 

 

 

 

 

display indication

description

measuring value

 

 

 

 

 

FREQ. (MHZ)

receiver frequency in MHz (numerical) and

MHz

 

 

bargraph center frequency indication in kHz

 

 

 

 

 

 

LEVEL (dBm)

receiver signal in dBm (numerical) and

dBm

 

 

bargraph indication (the lower bargraph

 

 

 

shows the set squelch threshold)

 

 

 

 

 

DISPL 1/2

AM-MOD.(90HZ)

AM modulation depth (90 Hz)

%

 

 

 

 

 

AM-MOD.(150HZ)

AM modulation depth (150 Hz)

%

 

 

 

 

 

DDM

differences in depth of modulation

DDM-value

 

 

 

(non-dimension value)

 

 

 

μA

 

 

 

%

 

 

 

 

 

SDM

total modulation factor

SDM-value

 

 

 

(non-dimension value)

 

 

 

μA

 

 

 

%

 

 

 

 

 

PHI (90/150)

phase shift (90 Hz/150 Hz)

degree

 

 

 

 

DISPL 2/2

ID AF-FREQ.

AF frequency (1020 Hz)

Hz

 

 

 

 

 

ID AM-MOD.

AM modulation depth of AF frequency

%

 

 

 

 

 

VOICE AM-MOD.

AM modulation depth in the range of

%

 

 

300 to 3000 Hz

 

 

 

 

 

Y / t Setup

Range 1 to 4

XY value scaling in localizer /

%

 

 

glideslope mode at DSP-Out.

 

 

 

 

 

STORE

Locations 1 to 120

DDM/SDM results

 

DDM menu

 

 

 

 

 

 

 

0796.1800.02