Procom EVS200 user manual Command AM2 Function, Command AM3 Function, Command PH Function

Models: EVS200

1 68
Download 68 pages 10.12 Kb
Page 46
Image 46

VOR/ILS Analyzer EVS200

 

 

Chapter 3: Operation

Command AM2

 

 

 

 

 

command

function

 

AM2 <AM modulation depth (90 Hz)>

query of AM modulation depth (90 Hz)

 

 

 

 

 

The command AM2 reads the actual AM modulation depth (90 Hz) in the ILS mode.

Example:

 

 

 

 

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

AM2 <CR>

e.g.

50% <CR/LF>

 

am2 <CR>

e.g.

50% <CR/LF>

Command AM3

command

function

 

AM3

<AM modulation depth (150 Hz)>

query of AM modulation depth (150 Hz)

 

The command AM3 reads the actual AM modulation depth (150 Hz) in the ILS mode.

Example:

 

 

 

command from the controller

response of the VOR/ILS Analyzer EVS200

Query:

AM3 <CR>

e.g.

29.9% <CR/LF>

 

am3 <CR>

e.g.

29.9% <CR/LF>

Command PH

command

function

 

PH

<Phase 90/150 Hz>

query of Phase angle 90/150 Hz

 

The command PH reads the actual PHI value between 90 Hz and 150 Hz in the ILS mode.

Example:

 

 

 

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

PH <CR>

e.g.

119.9DEG <CR/LF>

 

ph <CR>

e.g.

119.9DEG <CR/LF>

0796.1800.02

E-8

37