VOR/ILS Analyzer EVS200

 

 

Chapter 3: Operation

Command FA2

 

 

 

 

 

command

function

 

FA2 <Fast DDM measuring on>

Activating the Fast DDM measurement

 

 

 

 

 

The command FA2 switches on the fast DDM measurement. The measuring, used only for landing path measurement can only be activated through a controller.

The command FA2 performs the continuous measurement and output of the DDM, RF level, Mod- factor 90 Hz and Modfactor 150 Hz with a transmission speed of about 20 measurement value sets (DDM, RF level, AM90, AM150) per second. This conforms to a time interval of 50 ms. The stated RS-232-transmission times are related to 19.200 Baud.

The measurement values can be read in and output as a text file corresponding to the terminal pro- gram.

To reach 20 meas/s the baud rate must be set to 19200 baud in the Setup of the

VOR/ILS Analyzer EVS200. It must be attended the acceptance of the data transfering rate by the terminal program.

￿The measurement can be stopped through the command FA0 (refer to Command

FA0).

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

 

input:

FA2 <CR>

output form: (DDM <CR >, RF level <CR >,

 

fa2 <CR>

AM90 <CR >, AM 150 <CR >)

 

With the command FA2 the echo

 

 

 

mode is automatically switched

DDM: four-digit with indication (dimensionless)

 

OFF.

example: -1000 <CR > (-0.1000 DDM)

 

 

five-digit with indication (μA)

 

 

example: -09890 <CR > (-98.90 μA )

 

 

RF level: three-digit with indication (dBm)

 

 

example: -400 <CR > (-40.0dBm)

 

 

 

122 <CR > (12.2dBm)

 

 

Modulation depth:

three-digit (%)

 

 

example:

202 <CR > (20.2 %)

 

 

 

004 <CR > (0.4 %)

1/90Hz = 11.1msec.

N

N+1

N+2

N+3

N+4

 

N+5

 

N+6

 

N+7

N+8

 

N+9

 

 

N+10

DSP-Analysis

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DATA-Transfer

 

DDM, AM90, AM150,

 

RS232-

 

 

DATA-Transfer

 

DDM, AM90, AM150,

RS232-

 

DATA-

 

 

 

 

RF-Level - Calculation

 

Transfer

 

 

 

RF-Level - Calculation

Transfer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TXD-Signal

 

 

 

 

 

 

 

 

N

 

 

 

 

 

 

 

N+4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

fromEVS200

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

40 msec.

 

 

10 msec.

 

 

 

0796.1800.02

E-8

40

Page 49