VOR/ILS Analyzer EVS200
Chapter 3: Operation
Command RES
command | function | |
RES | <RESET> | |
| ||
With the command RES a unit master reset is possible. It works like a restart of the unit. | ||
Example: |
| |
| command from the controller | response of the VOR/ILS Analyzer EVS200 |
input: | RES<CR> | READY <CR/LF> |
| res<CR> | READY <CR/LF> |
EVS200
Command BI
command | function | |
BI |
| BITE (Build In Test) |
|
|
|
The EVS200 performs approx. every 500 ms a self test (BITE) and makes the result available as a 9bit binary form. During this test the following voltages and funtionally subunits of the units are checked.
all operating voltages,
hostprocessor
memory,
battery capacity (%)
explanation of BITE information
B8 |
| B7 | B6 |
| B5 | B4 | B3 | B2 | B1 | B0 |
SYNTH 1 | SYNTH 2 |
| +12 VDC | Batt2 | Batt1 | Batt0 | ||||
X |
| X | X |
| X | X | X | X | X | X |
X = 1 | is function in tolerance (ok) |
|
|
|
|
|
| |||
X = 0 | is function out of tolerance (failure) |
|
|
|
|
| ||||
Example: |
|
|
|
|
|
|
|
|
| |
|
| command from the controller | response of the VOR/ILS Analyzer EVS200 | |||||||
query: | BI <CR> |
|
| BI110111101 <CR/LF> |
|
0796.1800.02 | 33 |