VOR/ILS Analyzer EVS200
0796.1800.02 E-8 49
Chapter 3: Operation

Level-Mode

Command SR

command function
SR <Set REF> reference level setting
The command SR is for setting the actual receiver level as a reference level in the reference level
window "
level"-mode.
Example: command from the controller response of the VOR/ILS Analyzer EVS200
input: SR <CR>
sr <CR> READY <CR/LF>
READY <CR/LF>

Command LR

command function
LR <level reference> reference level query
The command LR reads the actual set reference level in the "
Level" mode.
Example: command from the controller response of the VOR/ILS Analyzer EVS200
query: LR <CR>
lr <CR> e.g. -48.1dBm <CR/LF>
e.g. -48.1dBm <CR/LF>

Command LD

command function
LR <level delta>
-Level query
The command LD reads the actual
-Level in the "
Level" mode.
Example: command from the controller response of the VOR/ILS Analyzer EVS200
query: LD <CR>
ld <CR> e.g. -000.0dBm <CR/LF>
e.g. -000.0dBm <CR/LF>