VOR/ILS Analyzer EVS200
|
| Chapter 3: Operation | |
Command FA1 |
|
|
|
|
| ||
command | function |
| |
FA1 <Fast DDM measuring on> | Activating the Fast DDM measurement |
| |
|
|
|
|
The command FA1 switches on the fast DDM measurement. The measuring, used only for landing path measurement can only be activated through a controller. The output is performed as DDM value with RF level (see example) and can be processed accordingly with calculation programs e.g. like Excel etc. A capacaty of 34 measurements /s at a baud rate set to 19200 baud will be reached. This conforms to a time interval of 30 ms. The measurement values (DDM, RF level) can be read in and output as a text file corresponding to the terminal program.
To reach 34 meas/s the baud rate must be set to 19200 baud in the Setup of the
VOR/ILS Analyzer EVS200. It must be attended the acceptance of the data transfering rate by the terminal program.
The measurement can be stopped through the command FA0 (refer to Command
FA0).
Example: | command from the controller | response of the VOR/ILS Analyzer EVS200 | |
| |||
input: | FA1 <CR> | 0.2008 | |
| fa1 <CR> | 0.2008 | |
| With the command FA1 the echo | Both values are separated through a tabulator | |
| mode is automatically switched |
| ASCII 9. |
| OFF. |
|
|
1/90Hz = 11.1msec.
N | N+1 | N+2 | N+3 | N+4 | N+5 | N+6 | N+7 | N+8 | N+9 | N+10 | |
| DATA- | DDM, | RS232- | DATA- | DDM, | RS232- | DATA- | DDM, | RS232- | DATA- |
|
| Transfer | Calculation | Transfer | Transfer | Calculation | Transfer | Transfer | Calculation | Transfer | Transfer |
|
|
|
| N |
|
| N+2 |
|
| N+5 |
| |
|
|
|
|
|
|
|
| fromEVS200 | |||
|
|
|
|
|
|
|
|
|
|
|
30 msec.
0796.1800.02 | 39 |