VOR/ILS Analyzer EVS200

 

 

Chapter 3: Operation

Command FA1

 

 

 

 

 

command

function

 

FA1 <Fast DDM measuring on>

Activating the Fast DDM measurement

 

 

 

 

 

The command FA1 switches on the fast DDM measurement. The measuring, used only for landing path measurement can only be activated through a controller. The output is performed as DDM value with RF level (see example) and can be processed accordingly with calculation programs e.g. like Excel etc. A capacaty of 34 measurements /s at a baud rate set to 19200 baud will be reached. This conforms to a time interval of 30 ms. The measurement values (DDM, RF level) can be read in and output as a text file corresponding to the terminal program.

To reach 34 meas/s the baud rate must be set to 19200 baud in the Setup of the

VOR/ILS Analyzer EVS200. It must be attended the acceptance of the data transfering rate by the terminal program.

￿The measurement can be stopped through the command FA0 (refer to Command

FA0).

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

 

input:

FA1 <CR>

0.2008

-48.2 <CR/LF>

 

fa1 <CR>

0.2008

-48.2 <CR/LF>

 

With the command FA1 the echo

Both values are separated through a tabulator

 

mode is automatically switched

 

ASCII 9.

 

OFF.

 

 

1/90Hz = 11.1msec.

N

N+1

N+2

N+3

N+4

N+5

N+6

N+7

N+8

N+9

N+10

DSP-Analysis

 

DATA-

DDM, RF-Level

RS232-

DATA-

DDM, RF-Level

RS232-

DATA-

DDM, RF-Level

RS232-

DATA-

 

 

Transfer

Calculation

Transfer

Transfer

Calculation

Transfer

Transfer

Calculation

Transfer

Transfer

 

 

 

 

N

 

 

N+2

 

 

N+5

 

TXD-Signal

 

 

 

 

 

 

 

 

fromEVS200

 

 

 

 

 

 

 

 

 

 

 

30 msec.

0796.1800.02

E-8

39

Page 48