•If you cannot execute an ADJUST or ANALYZE menu operation when using the Sony
•When using the Thoma TF6 probe, set the TF6 PRINT menu to off.
[B121] SONY
ADJUST... : Perform automatic color temperature adjustment. ⎟ [B1211]
ANALYZE: Display readout values on the screen.
⎟[B1212]
[B1211] ADJUST menu
To start adjustment, proceed as follows.
When you use the previously adjusted values for adjustment, you can make start adjustment by selecting START without performing step (1) and step (2).
(1)Select either D65 or D93.
Rather than selecting D65 or D93, you may instead enter the values of the CIE 1931 color system x and y coordinates.
(2)Enter values for LOW LIGHT and HIGH LIGHT.
(3)Select START.
START: Start adjustment. ⎟ [B12111]
X:Enter the x coordinate.
Y:Enter the y coordinate.
LOW LIGHT (20IRE): Enter the luminance (cd/m2) for low light.
HIGH LIGHT (100IRE): Enter the luminance (cd/ m2) for high light.
D65: Use D65 setting.
D93: Use D93 setting.
[B12111] START menu
The following message appears. Perform operation according to the message to start adjustment.
SET PROBE ON CRT
PRESS ENTER
Adjustment starts when the probe is placed against the center of the screen and the ENTER or Ent button is pressed.
SET PROBE ON THIS CRT
PRESS ENTER
Attach the
[B12121] SET PROBE ON THIS CRT menu
Display the color system and luminance readout values from the
X:xxxx: Display the x coordinate of the color system.
Y:xxxx: Display the y coordinate of the color system.
L:xxxx: Display the luminance value.
[B122] GRASEBY/MINOLTA/PHILIPS/THOMA menu
The forth and lower levels of [B122] are the same as those of the fifth and lower levels of [B1211]. Refer to [B1211] and [B12111].
[B13] COPY FROM menu
Select the source to be copied from.
OTHER VALUE... : Copy data from another channel or from common data. ⎟ [B131]
OTHER MONITOR... : Copy data from a BVM- DxxE/DxxF series monitor. ⎟ [B132]
MEMORY CARD... : Copy data from a monitor memory card. ⎟ [B133]
[B131] OTHER VALUE menu
Select STD, COL1,COL2, or CH SET. ⎟ Copy is carried out.
STD: Copy common data (factory setting: D65).
COL1: Copy common data (factory setting: D93).
COL2: Copy common data (factory setting: D56/ D651)).
CH SET: Copy data from a particular channel (factory setting: D65). Enter the number of the channel from which the data will be copied.
D65: Copy the color temperature of D65.
D93: Copy the color temperature of D93.
[B132] OTHER MONITOR menu Specify the address number of the monitor.
2 Chapter Menu
[B1212] ANALYZE menu | MONITOR ADDRESS: Input the address number of |
The following message appears. Perform operation | the monitor from which the data will be copied. |
according to the message to enable the | ⎟ [B1321] |
read the color system and luminance value.
........................................................................................................................................................................................................
1)D56:
D65:
33