Suburban Mfg 12019-01-NGFREP manual Knob and Lock Screw

Models: 12019-01-NGFREP

1 184
Download 184 pages 31.08 Kb
Page 107
Image 107

Knob

P/N 853461

Owner’s Manual - Regal Intensive Care Unit

Three knobs are used on the electronics compartment door. One is on the oxygen concentration meter calibration potentiometer, and the others are the left and right floor temperature controller knobs. The knobs are identical and are removed and installed the same way.

Tool Required

„Small flat-blade screwdriver

Procedure

1.Take note of the position of the indicator line on the knob (Figure 107).

2.With a small flat-blade screwdriver, loosen the lock screw on the side of the knob.

3.Pull the knob off the shaft.

4.Press the new knob onto the shaft with the indicator line in approximately the same position as on the old knob.

5.Tighten the lock screw.

Indicator Line

Lock

Screw

Figure 107. Knob and Lock Screw

Form No. 702718 Rev. A March, 2005

95

Page 107
Image 107
Suburban Mfg 12019-01-NGFREP manual Knob and Lock Screw