Suburban Mfg 12019-01-NGFREP manual Excessive carbon dioxide in the animal compartment

Models: 12019-01-NGFREP

1 184
Download 184 pages 31.08 Kb
Page 163
Image 163

Owner’s Manual - Regal Intensive Care Unit

Excessive carbon dioxide in the animal compartment.

Remedial Action

First: The Sodalime supply in the filter/fans compartment is probably exhausted. Replace the Sodalime with fresh material. Refer to Refilling the CO2 Filter on Page 65.

Second: Make sure the circulating fans in the filter/fans compartment are working. Refer to Circulating fan does not work on Page 149.

Third: Make sure that the air inlet grill and the air outlet grill (Figure 3 on Page 9) are clear and unobstructed.

Fourth: Make sure that your set-point setting on the oxygen concentration meter is correct. Refer to Using Oxygen on Page 51.

Fifth: Make sure that the flow rate on the oxygen flow meter is correct. Refer to Controlling Oxygen Flow Rate on Page 56.

Sixth: Check your oxygen supply and pressure regulator setting.

Seventh: Refer to Oxygen flow is inadequate on Page 141.