Suburban Mfg 12019-01-NGFREP manual Circulating fan does not work

Models: 12019-01-NGFREP

1 184
Download 184 pages 31.08 Kb
Page 161
Image 161

Owner’s Manual - Regal Intensive Care Unit

Circulating fan does not work.

Remedial Action

First: Make sure that electrical power to the unit is OK and that the main on/off switch is On.

Second: Check the circuit breakers to be sure they haven’t tripped. Reset them if necessary. Refer to Re-setting the Circuit Breakers on Page 60.

CAUTION: Before working on the electrical system, make sure the unit is Off and the electric power cord un-plugged.

Third: Remove the rear utility panel (Figure 139 on Page 123) and check the wire connections to the fans in the rear of the electronics compartment (Figures 158 and 159 on Page 131). Make sure the connections are tight. If the ICU fan wiring has been serviced lately, make sure that the wires have been re-connected correctly.

Fourth: Are all three circulating fans not working, or just one? Is the electronics compartment ventilating fan working? If one circulating fan is not working but the the other two are OK, check the power cord connection on that one fan. If the power cord is secure, you probably have a bad fan. Replace the fan. Order P/N 853414 and refer to Page 134 for replacement instructions.

Fifth: If the other circulating fans are not working either, check the other electrical functions in the ICU. If none of the electrical func- tions are working, refer to The ICU will not turn on on Page 139.

Form No. 702718 Rev. A March, 2005