7.10.6

Status Control Commands

 

Refer to Section 7-8 for definition of the registers.

 

#

 

Command

 

Description

1

 

STT?

Reads the complete power supply status.

 

 

 

Returns ASCII characters representing the following data, separated by commas:

 

 

 

MV<actual (measured) voltage>

PC<programmed (set) current>

 

 

 

PV<programmed (set) voltage>

SR<status register, 2-digit hex>

 

 

 

MC<actual (measured) current>

FR<fault register, 2-digit hex>

 

 

 

Example response: MV(45.201),PV(45), MC(4.3257), PC(10), SR(30), FR(00)

2

 

FLT?

Reads Fault Conditional Register. Returns 2-digit hex.

3

 

FENA

Set Fault Enable Register using 2-digit hex.

4

 

FENA?

Reads Fault Enable Register. Returns 2-digit hex.

5

 

FEVE?

Reads Fault Event Register. Returns 2-digit hex. Clears bits of Fault Event Register.

6

 

STAT?

Reads Status Conditional Register. Returns 2-digit hex.

7

 

SENA

Sets Status Enable Register using 2-digit hex.

8

 

SENA?

Reads Status Enable Register. Returns 2-digit hex.

9

 

SEVE?

Reads Status Event register. Returns 2-digit hex. Clears bits of Status Event register.

7.11 STATUS, ERROR AND SRQ REGISTERS

7.11.1 General Description

This Section describes the various status error and SRQ registers structure. The registers can be read or set via the RS232/RS485 commands. When using the IEEE option, refer to the User’s Manual for GenesysTM Power Supply IEEE Programming Interface.

Refer to Fig. 7-7 for the Status and Error Registers Diagram.

62

83-507-013 Rev. D

Page 70