CHAPTER 6 REMOTE ANALOG PROGRAMMING
6.1 INTRODUCTION
The rear panel connector J1 allows the user to program the power supply Output Voltage and Current with an analog device. J1 also provides monitoring signals for Output Voltage and Output Current. The programming range and monitoring signals range can be selected between
CAUTION
COM
6.2 LOCAL/REMOTE ANALOG INDICATION
Contact 8 of J1
In Local mode, the Output Voltage and Output Current can be programmed via the front panel VOLTAGE and CURRENT encoders or via the RS232/RS485 port. In Remote Analog mode, the Out- put Voltage and current can be programmed by analog voltage or by programming resistors via J1 contacts 9 and 10 (refer to Sections 6.4 and 6.5). Refer to Table
Output Voltage/ | |||
Current setting | |||
|
| ||
Down (default) | No effect | Local | |
Up | “0” or Short | Remote | |
“1” or Open | Local | ||
|
Table
6.3 LOCAL/REMOTE ANALOG INDICATION
Contact 21 of J1 (Fig.
Mode | ||||
TTL “0” or short | Down | Down | Open | Local (FP) |
| Down | Up | 00.6V | Remote Analog |
| Up | Down | 00.6V | Remote Analog |
| Up | Up | 00.6V | Remote Analog |
TTL “1” or open | Down or Up | Down or Up | Open | Local (FP) |
Table