Total Organic Carbon Analyzer

Maintenance

 

 

CAUTION: THE UV LIGHTED LENGTH MUST BE AT LEAST 12 ½ INCHES FOR PROPER OPERATION. DO NOT USE STANDARD 12 INCH LIGHTED LENGTH UV LAMPS.

2.Turn off power allow 10 minutes for cool down.

3.Disconnect electrical connectors.

4.Loosen fittings as illustrated.

5.Remove UV Lamp.

6.Recalibrate analyzer.

5.3.7UV Reactor Assembly (P/N ST20003-1)

If malfunctioning, clean the assembly. If still malfunctioning, replace the assembly (P/N ST20003-1). See Figure 5-9.

Figure 5-9: Removing the UV Reactor Assembly

To Replace the UV Reactor Assembly:

1.Turn off UV Lamp at the UV Power Supply. Remove Persulfate and Acid Reagent lines from containers, as

Teledyne Analytical Instruments

57

Page 69
Image 69
Teledyne 6750 operating instructions UV Reactor Assembly P/N ST20003-1, Removing the UV Reactor Assembly