Teledyne 6750 operating instructions 4.3Detailed Calibration Procedures

Models: 6750

1 96
Download 96 pages 40.67 Kb
Page 48
Image 48
4.3Detailed Calibration Procedures

Setup and Operation

Model 6750

 

 

The following window is illustrative of the D/DBPR [Run] screen.

4.3Detailed Calibration Procedures

1.Although the instrument has been calibrated to your specifications at the factory, it should be rechecked after satisfactory installation and periodically calibrated as suggested for your application.

2.Instrument calibration is performed by an “end-to-end” method, whereby a known organic carbon chemical standard solution is introduced to the analyzer and the analyzer is “spanned” to that value.

3.Organic and Inorganic Carbon Standards:

a.(Potassium Hydrogen Phthalate) is recommended for the organic carbon standard solution.

b.Sodium Carbonate is recommended for the inorganic carbon standard.

c.Table 4-1 provides the concentration to be used for two ranges of different chemical compounds (organic & inorganic carbon).

d.Ratioing of these concentrations will provide other ranges for the span solution.

Teledyne Analytical Instruments

36

Page 48
Image 48
Teledyne 6750 operating instructions 4.3Detailed Calibration Procedures