Table 4-1: Preparation of Standards (TOC/TIC)

Total Organic Carbon Analyzer

Setup and Operation

 

 

 

Table 4-1: Preparation of Standards (TOC/TIC)

 

ORGANIC COMPOUNDS

 

 

100 mg/Liter

 

 

1000 mg/Liter

 

 

 

 

TOC (AS CARBON)

 

 

TOC (AS CARBON)

 

 

 

 

 

 

 

 

 

Ethylene Glycol

 

.233 ml/L H20

 

2.33 ml/L H20

 

Methanol

 

.337 ml/L H20

 

3.37 ml/L H20

 

Ethanol

 

.242 ml/L H20

 

2.42 ml/L H20

 

Acetone

 

.204 ml/L H20

 

2.04 ml/L H20

 

Carbon Tetrachloride

 

.807 ml/L H20

 

8.07 ml/L H20

 

Sucrose

 

.238 gm/L H20

 

2.38 gm/L H20

 

Urea

 

.500 gm/L H20

 

5.00 gm/L H20

 

Acetic Acid

 

.250 gm/L H20

 

2.50 gm/L H20

 

KHP (Potassium acid

 

.212 gm/L H20

 

2.12 gm/L H20

 

phthalate)

 

 

 

 

 

 

 

 

 

 

Glycine

 

.313 gm/L H20

 

3.13 gm/L H20

 

Sodium Stearate

 

.141 gm/L H20

 

1.41 gm/L H20

 

Succinic Acid

 

.246 gm/L H20

 

2.46 gm/L H20

 

Sodium Oxalate

 

.555 gm/L H20

 

5.55 gm/L H20

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

INORGANIC COMPOUNDS

 

 

50 mg/Liter

 

 

 

 

TIC (AS CARBON)

 

 

 

 

 

 

Sodium Carbonate

.442 gm/L H20

Potassium Carbonate

.575 gm/L H20

Ammonium Carbonate

.475 gm/L H20

Add amounts shown (in milliliters or grams) to clean dry, one liter volumetric flask and dilute to one (1) liter with distilled, deionized water.