Maintenance

Model 6750

 

 

Figure 5-11: Mass Flow Controller

5.3.10 Metering Valve (P/N ST16050)

If no O2/air flow to the sparger (P/N ST20025) is observed (no bubbles), check for tubing and fitting leaks, restrictions and adjust metering valve (be careful to return to previous indicated bubble rate). If flow cannot be restored, replace the metering valve (P/N ST16050) according to the directions below. See Figure 5-12.

To replace the metering valve:

1.Turn carrier gas (Air/O2) OFF.

2.Turn pumps OFF.

3.Remove fittings and tubing.

4.Remove metering valve.

5.Recalibrate TIC if TIC and/or TOC–TRUE analysis is required. NPOC analysis only does not require recalibration.

6.Reinstall metering valve in reverse order.

Teledyne Analytical Instruments

60

Page 72
Image 72
Teledyne 6750 operating instructions Mass Flow Controller, Metering Valve P/N ST16050