Maintenance

Model 6750

 

 

5.3.4 D. C. Power Supply

The power supply has no field adjustment. If malfunctioning, replace the module (P/N ST15083) using the directions below. See Figure 5-6.

Figure 5-6: DC Power Supply Module

To replace the module:

1.Turn off power.

2.Remove plastic cover.

3.Disconnect connectors.

4.Remove nuts holding Power Supply.

5.Remove Power Supply Module.

6.Replace Power Supply Module.

7.Reverse steps for installation.

8.(Refer to Installation Drawings for proper connector assembly.)

9.Recalibrate liquid and span values.

Teledyne Analytical Instruments

54

Page 66
Image 66
Teledyne 6750 operating instructions 4 D. C. Power Supply, DC Power Supply Module