Teledyne 6750 operating instructions Setup and Operation, 4.1Analyzer Startup

Models: 6750

1 96
Download 96 pages 40.67 Kb
Page 41
Image 41
Setup and Operation

Total Organic Carbon Analyzer

Setup and Operation

 

 

 

Setup and Operation

Once the analyzer has been installed, it can be configured to your application. To do this you will:

Set system parameters

Calibrate the instrument

Define the analysis range. Then choose auto-ranging or select a fixed range of analysis, as required.

Set alarm setpoints, and modes of alarm operation (latching, failsafe, etc.)

4.1Analyzer Startup

Referring to the Installation section and “AS-BUILT” Drawings in the Appendix, check the following:

1.Air/oxygen supply – This must provide a constant

pressure of 15 +/- 2.0 psi, ultra-pure oxygen or CO2–free air at a flow rate up to 500 cc/minute. (Pre-purified air may be used with optional Air Purifier P/N ST910003).

2.Drain Line – Check to see if the drain line is free of kinks and loops and allows normal vented gravity flow to an open receptacle (jar, pipe, etc.). AN AIR GAP AS DEFINED IS MANDATORY.

CAUTION: FAILURE TO OBSERVE THIS WILL RESULT IN IMPROPER OPERATION.

3.ELECTRICAL – Assure proper electrical installation. Refer to facility requirements.

4.SAFETY – All customer and Teledyne specified safety measures are followed.

Teledyne Analytical Instruments

29

Page 41
Image 41
Teledyne 6750 operating instructions Setup and Operation, 4.1Analyzer Startup