4.2.4 Stream Sequencer

Setup and Operation

Model 6750

 

 

From this menu verify or change the concentration alarm setpoints using the onscreen scroll arrows.

4.2.4 Stream Sequencer

If your analyzer has been configured for a multi-stream sequencer (P/N ST200009), from the [Edit] Menu, select [Sequencer]. The following menu appears:

From this menu, you can:

Configure the duration of analysis for each stream (how long it will analyze that stream before selecting the next stream)

Configure the changeover time (to allow the prior stream to pass through sampling systems and clear the analyzer before analysis of the next steam begins),

Enable and disable each stream analysis (for example, to concentrate on only one stream).

The following window is illustrative of a dual stream [Run] screen.