Total Organic Carbon Analyzer | Installation | |
|
|
|
With the exception of the vent and carrier gas inlet, all other gas connections are made on the same bottom panel as the liquid connections and are shown in Figure
Gas Carrier Connections:
The facility carrier gas must be free of CO2 and hydrocarbons and be regulated to 15±2 psi and connected as shown in “As Built” Drawings in the Appendix).
NDIR Vent (gas exhaust):
Exhaust connections must be consistent with the hazard level of the constituent gases. Check Local, State, and Federal laws, and ensure that the exhaust stream vents to an appropriately controlled area if required.
CAUTION: OPERATING THE UNIT WITHOUT CARRIER GAS
CAN DAMAGE THE ANALYZER.
3.5 Checking the System
Before applying power to the instrument, check:
•All gas and liquid lines as well as vent and drains are properly connected and there are no leaks in the lines or at fittings.
•Power and signal wiring are correct and that there are no frayed or loose wires that could cause a short.
Prior to powering up the system and using the instrument for analysis, the system must be configured for your application and calibrated. This is described in Section 4.
Teledyne Analytical Instruments | 27 |